Navien · General troubleshooting
Replace the TDS sensor for the product/regeneration water pipe
What is going on
The CIP tank must come out to reach this sensor. Be careful with CIP solution: a spill can corrode the unit, and if it contacts skin or eyes, flush with running water right away.
Check in this order
- Turn off the feedwater supply.
- Disconnect the power supply from the WEC.
- Remove the 2 screws on each side of the front cover and set the cover aside.
- Remove the 4 screws from the CIP tank assembly and pull the tank out of the WEC. Avoid spilling the solution.
- Place the CIP tank on a flat, clear surface.
- From the back of the CIP tank, disconnect the water level sensor connector and CIP dosing pump connector from the harness.
- Disconnect the TDS sensor connector from the harness.
- Loosen the sensor nut with a wrench and remove the TDS sensor assembly from the Y-adapter.
- Install the new TDS sensor assembly on the Y-adapter and tighten the nut with a wrench.
- Connect the TDS sensor connector to the harness.
- From the back of the CIP tank, reconnect the water level sensor and CIP dosing pump connectors to the harness.
- Install the CIP tank in the WEC and secure with 4 screws.
- Align the front cover with the WEC and secure each side with 2 screws.
Applies to: WEC600
More general guides
Take it to the job
All guides, error codes and tips work offline in the HVACopedia app.
Source: Navien WEC service manual, p. 97-98. Written in our own words; check the manual for your exact model.
For licensed technicians. Gas, electrical and refrigerant work must be done by a qualified, licensed person. Shut off power and gas before servicing, never bypass a safety control, and follow the manufacturer’s manual for the exact model; it and local codes always take priority. HVACopedia is not affiliated with any manufacturer; brand and model names only identify the equipment.