Navien · General troubleshooting

Removing the NDI modules

WEC Series

What is going on

Removes the NDI modules by taking off the covers, top frame, flow sensor housing and the product/regeneration water pipe assembly. Remove the CIP tank for access where needed.

Check in this order

  1. Shut off the feedwater supply.
  2. Disconnect power to the WEC.
  3. Remove the front and top covers (steps 3-6 in 8.2 Replacing the Power Assembly, p. 83).
  4. Remove the back cover (see Removing the Back Cover, p. 82).
  5. Remove 6 screws from the front side of the WEC.
  6. Remove 4 screws from the back side of the WEC and remove both side covers.
  7. Unplug the outlet water pipe, product/regeneration water pipe and NDI module connectors from the harness.
  8. Unplug the circulation pump connector from the harness.
  9. Remove the 3 plastic pins and 2 plastic pins that hold the harness to the top frame.
  10. Remove the 4 top frame screws, 2 front screws and 4 side screws (2 per side), then lift out the top frame assembly.
  11. Remove 4 screws that hold the flow sensor housing to the NDI modules.
  12. Remove 4 clips from the Y-adapters and the circulation pump outlet hose, detach the hose, and remove the product/regeneration water pipe assembly.
  13. Remove 4 screws from the side frame brackets.
  14. Remove 4 screws from the CIP tank, remove 2 clips from the CIP dosing pump, and lift the CIP tank assembly out of the WEC.
  15. Unplug the water level sensor and CIP dosing pump connectors from the harness and detach the CIP outlet hose from the dosing pump.

Applies to: WEC600

More general guides

Take it to the job

All guides, error codes and tips work offline in the HVACopedia app.

Open the app

Source: Navien WEC Series service manual, p. 110-112. Written in our own words; check the manual for your exact model.

For licensed technicians. Gas, electrical and refrigerant work must be done by a qualified, licensed person. Shut off power and gas before servicing, never bypass a safety control, and follow the manufacturer’s manual for the exact model; it and local codes always take priority. HVACopedia is not affiliated with any manufacturer; brand and model names only identify the equipment.